| 1 |
Device Identification Record manufacturer, model (LIFEPAK 1000), serial number, asset ID, firmware/software version, battery model/lot. |
|
|
|
| 2 |
Visual Inspection Inspect casing, screen, buttons, speaker, connectors, battery compartment and labels. No damage, corrosion or fluid ingress. |
|
|
|
| 3 |
Accessories Present Confirm adult pads present and in date. Confirm paediatric pads/attenuator if supplied. Confirm battery present. |
|
|
|
| 4 |
Consumables Status Confirm pads and battery are within service life. Replace via Service Action if required (do not fail unless functional impact). |
|
|
|
| 5 |
Engineering Setup Remove device from carry case. Disconnect pads. Prepare electrical safety analyser and defibrillator analyser. |
|
|
|
| 6 |
Power State Declaration Select power state used for testing. Mains connection triggers full electrical safety testing. |
|
|
|
| 7 |
Protective Earth Resistance (Mains only) Measure protective earth resistance from earth pin to exposed metal. |
|
|
|
| 8 |
Enclosure Leakage Current (NC) (Mains only) Measure enclosure/touch leakage current under normal condition. |
|
|
|
| 9 |
Patient Leakage Current – ECG (NC) (If ECG fitted) Measure patient leakage via ECG applied parts (BF). |
|
|
|
| 10 |
Defibrillator Analyser Setup Connect LIFEPAK 1000 to calibrated defibrillator analyser set to 50 ohm load using test cable or pads. |
|
|
|
| 11 |
Power-On Self Test Power on device. Confirm no fault codes and readiness indicator OK. |
|
|
|
| 12 |
Speaker and Prompts Confirm voice prompts and tones are clear and audible. |
|
|
|
| 13 |
Charge Time @ Maximum Energy Select maximum adult energy and measure charge time to ready-to-shock. |
|
|
|
| 14 |
Delivered Energy @ 200J (50Ω) Deliver 200J shock into 50Ω and record delivered energy. |
|
|
|
| 15 |
Delivered Energy @ 300J (50Ω) Deliver 300J shock into 50Ω and record delivered energy. |
|
|
|
| 16 |
Delivered Energy @ 360J (50Ω) Deliver 360J shock into 50Ω and record delivered energy. |
|
|
|
| 17 |
Post-Shock Recovery Confirm device returns to analysis/ready state without error after shock. |
|
|
|
| 18 |
Rhythm Simulator Setup Prepare rhythm simulator for VF, pVT, Asystole and NSR waveforms. |
|
|
|
| 19 |
Rhythm Analysis – VF Simulate ventricular fibrillation. Confirm 'Shock Advised'. |
|
|
|
| 20 |
Rhythm Analysis – pVT Simulate pulseless VT. Confirm 'Shock Advised'. |
|
|
|
| 21 |
Rhythm Analysis – Asystole Simulate asystole. Confirm 'No Shock Advised'. |
|
|
|
| 22 |
Rhythm Analysis – NSR Simulate normal sinus rhythm. Confirm 'No Shock Advised'. |
|
|
|
| 23 |
CPR Prompt Interval Measure CPR interval following shock or no-shock decision. |
|
|
|
| 24 |
Paediatric Setup Fit Infant/Child pads or attenuator if applicable. |
|
|
|
| 25 |
Paediatric Energy @ 200J Selected With paediatric pads fitted, select 200J and measure delivered energy. |
|
|
|
| 26 |
Paediatric Energy Cap Verification Select maximum adult energy and confirm paediatric output does not exceed 86J nominal. |
|
|
|
| 27 |
Configuration – Stacked Shocks Disabled Verify stacked shocks are OFF (single shock then CPR). |
|
|
|
| 28 |
Configuration – CPR Time After Shock Verify CPR interval after shock is 120 seconds. |
|
|
|
| 29 |
Configuration – CPR Time After No Shock Verify CPR interval after no-shock is 120 seconds. |
|
|
|
| 30 |
Configuration – Pulse Check Prompt Verify pulse check prompt behaviour. |
|
|
|
| 31 |
Final Engineering Check Reconnect pads, refit battery, clean device, ensure correct reassembly. |
|
|
|
| 32 |
Ready for Clinical Use Confirm device operational, labelled, and safe for return to service. |
|
|
|
| 33 |
Technician Comments Record advisories, service actions performed, parts replaced, or deviations. |
|
|
|