Test Run Details

Test Run Information
Export PDF InProgress
Lifepak 1000 - Annual Test
David Bradley
2026-01-30 01:35
Test run is paused. Resume the test run before completing it.
# Step Type Result Pass/Fail Comment
1 Device Identification
Record manufacturer, model (LIFEPAK 1000), serial number, asset ID, firmware/software version, battery model/lot.
IEC 62353
Text Note - - -
2 Visual Inspection
Inspect casing, screen, buttons, speaker, connectors, battery compartment and labels. No damage, corrosion or fluid ingress.
IEC 62353 §4.2
Yes/No Check - - -
3 Accessories Present
Confirm adult pads present and in date. Confirm paediatric pads/attenuator if supplied. Confirm battery present.
Manufacturer
Yes/No Check - - -
4 Consumables Status
Confirm pads and battery are within service life. Replace via Service Action if required (do not fail unless functional impact).
Service Policy
Yes/No Check - - -
5 Engineering Setup
Remove device from carry case. Disconnect pads. Prepare electrical safety analyser and defibrillator analyser.
Engineering
Instruction - - -
6 Power State Declaration
Select power state used for testing. Mains connection triggers full electrical safety testing.
Expected:{"Value":"Mains connected"}}"
Multiple Choice - - -
7 Protective Earth Resistance
(Mains only) Measure protective earth resistance from earth pin to exposed metal.
IEC 62353 / IEC 60601-1
Measurement - - -
8 Enclosure Leakage Current (NC)
(Mains only) Measure enclosure/touch leakage current under normal condition.
IEC 62353
Measurement - - -
9 Patient Leakage Current – ECG (NC)
(If ECG fitted) Measure patient leakage via ECG applied parts (BF).
IEC 62353
Measurement - - -
10 Defibrillator Analyser Setup
Connect LIFEPAK 1000 to calibrated defibrillator analyser set to 50 ohm load using test cable or pads.
IEC 60601-2-4
Instruction - - -
11 Power-On Self Test
Power on device. Confirm no fault codes and readiness indicator OK.
Manufacturer
Yes/No Check - - -
12 Speaker and Prompts
Confirm voice prompts and tones are clear and audible.
Manufacturer
Yes/No Check - - -
13 Charge Time @ Maximum Energy
Select maximum adult energy and measure charge time to ready-to-shock.
IEC 60601-2-4
Measurement - - -
14 Delivered Energy @ 200J (50Ω)
Deliver 200J shock into 50Ω and record delivered energy.
J
Measurement - - -
15 Delivered Energy @ 300J (50Ω)
Deliver 300J shock into 50Ω and record delivered energy.
J
Measurement - - -
16 Delivered Energy @ 360J (50Ω)
Deliver 360J shock into 50Ω and record delivered energy.
J
Measurement - - -
17 Post-Shock Recovery
Confirm device returns to analysis/ready state without error after shock.
IEC 60601-2-4
Yes/No Check - - -
18 Rhythm Simulator Setup
Prepare rhythm simulator for VF, pVT, Asystole and NSR waveforms.
IEC 60601-2-4
Instruction - - -
19 Rhythm Analysis – VF
Simulate ventricular fibrillation. Confirm 'Shock Advised'.
IEC 60601-2-4
Yes/No Check - - -
20 Rhythm Analysis – pVT
Simulate pulseless VT. Confirm 'Shock Advised'.
IEC 60601-2-4
Yes/No Check - - -
21 Rhythm Analysis – Asystole
Simulate asystole. Confirm 'No Shock Advised'.
IEC 60601-2-4
Yes/No Check - - -
22 Rhythm Analysis – NSR
Simulate normal sinus rhythm. Confirm 'No Shock Advised'.
IEC 60601-2-4
Yes/No Check - - -
23 CPR Prompt Interval
Measure CPR interval following shock or no-shock decision.
seconds
Measurement - - -
24 Paediatric Setup
Fit Infant/Child pads or attenuator if applicable.
IEC 60601-2-4
Instruction - - -
25 Paediatric Energy @ 200J Selected
With paediatric pads fitted, select 200J and measure delivered energy.
Limits:{"Max":86}}"
Measurement - - -
26 Paediatric Energy Cap Verification
Select maximum adult energy and confirm paediatric output does not exceed 86J nominal.
J
Measurement - - -
27 Configuration – Stacked Shocks Disabled
Verify stacked shocks are OFF (single shock then CPR).
AHA / Manufacturer
Yes/No Check - - -
28 Configuration – CPR Time After Shock
Verify CPR interval after shock is 120 seconds.
seconds
Measurement - - -
29 Configuration – CPR Time After No Shock
Verify CPR interval after no-shock is 120 seconds.
seconds
Measurement - - -
30 Configuration – Pulse Check Prompt
Verify pulse check prompt behaviour.
ALWAYS]
Multiple Choice - - -
31 Final Engineering Check
Reconnect pads, refit battery, clean device, ensure correct reassembly.
Engineering
Instruction - - -
32 Ready for Clinical Use
Confirm device operational, labelled, and safe for return to service.
Service Release
Yes/No Check - - -
33 Technician Comments
Record advisories, service actions performed, parts replaced, or deviations.
Quality Record
Text Note - - -
Back to List Edit Results
Admin Customizer

Easily configure layout, styles, and preferences for your admin interface.

Select Theme
Default
Material
Modern
SaaS
Flat
Minimal
Color Scheme
Light
Dark
System
Topbar Color
Light
Dark
Gray
Gradient
Sidenav Color
Light
Dark
Gray
Gradient
Image
Sidebar Size
Default
Compact
Condensed
On Hover
On Hover - Show
Offcanvas
Layout Position