Template Steps

Medical Electrical Safety – IEC 62353 (Recurrent)
v1.0  |  2025-10-02 00:07Published
Test Steps
# Title Type Limits Unit Report Required Actions
1
Protective Earth Resistance
Measure PE from earth pin to accessible conductive parts.
NumericMeasure - Ω Locked
2
Enclosure Leakage Current (NC)
Measure enclosure leakage in normal condition using DIRECT or DIFFERENTIAL method.
NumericMeasure - µA Locked
3
Patient/Applied-Part Leakage (NC)
Measure patient leakage in normal condition.
NumericMeasure - µA Locked
4
Patient/Applied-Part Leakage (SFC, open earth)
Simulate open protective earth and measure leakage.
NumericMeasure - µA Locked
5
Insulation Resistance
Apply 500 VDC between mains parts and protective earth.
NumericMeasure - Locked