Template Steps

Lifepak 1000 - Annual Test
v1  |  2026-01-30 10:05Draft
Test Steps
# Title Type Limits Unit Report Required Actions
1
Visual Inspection
Inspect casing, screen, buttons, speaker, connectors, battery compartment and labels. No damage, corrosion or fluid ingress.
BooleanCheck - -
2
Accessories Present
Confirm adult pads present and in date. Confirm paediatric pads/attenuator if supplied. Confirm battery present.
BooleanCheck - -
3
Consumables Status
Confirm pads and battery are within service life. Replace via Service Action if required (do not fail unless functional impact).
BooleanCheck - -
4
Battery Serial Number
TextNote - -
5
Battery Expiry Date
TextNote - -
6
Speaker and Prompts
Confirm voice prompts and tones are clear and audible.
BooleanCheck - -
7
Service Menu
Disconnect pads. Prepare electrical safety analyser and defibrillator analyser.
Instruction - -
8
Confirm correct mahine Date & Time
BooleanCheck - -
9
Machine Pixel Test
Run machine pixel test
BooleanCheck - -
10
Power-On Self Test
Power on device. Confirm no fault codes and readiness indicator OK.
BooleanCheck - -
11
Configuration: Energy Levels
E1: 200, E2: 300, E3: 360
BooleanCheck - -
12
Configuration: Delete after send
Expected: False
BooleanCheck - -
13
Configuration: CPR
CPR T1: 120, CPR T2: 120, Initial CPR: Off, Preshock: Off, CPR Prompt: On
BooleanCheck - -
14
Configuration – CPR Time After Shock
Verify CPR interval after shock is 120 seconds.
NumericMeasure - -
15
Configuration: Pulse Check
Expected: False
BooleanCheck - -
16
Configuration: Manual Mode
Expected: Off
BooleanCheck - -
17
Configuration – Stacked Shocks Disabled
Verify stacked shocks are OFF (single shock then CPR).
BooleanCheck - -
18
Delivered Energy @ 100J (50Ω)
Deliver 100J shock into 50Ω and record delivered energy.
NumericMeasure 85 - 115 J J
19
Delivered Energy @ 360J (50Ω)
Deliver 360J shock into 50Ω and record delivered energy.
NumericMeasure 306 - 414 J J
20
Defibrillator Analyser Setup
Connect LIFEPAK 1000 to calibrated defibrillator analyser set to 50 ohm load using test cable or pads.
Instruction - -
21
ECG: 60bpm NSR
NumericMeasure 56 - 64 bpm bpm
22
ECG: 120bpm NSR
NumericMeasure 116 - 124 bpm bpm
23
ECG: 180bpm NSR
NumericMeasure 176 - 184 bpm bpm
24
ECG: VF -> Defib Prompt
Machine prompts user to connect therapy cable when VF detected in ECG mode
BooleanCheck - -
25
Technician Comments
Record advisories, service actions performed, parts replaced, or deviations.
TextNote - -
26
Final Engineering Check
Reconnect pads, refit battery, clean device, ensure correct reassembly.
Instruction - -
27
Ready for Clinical Use
Confirm device operational, labelled, and safe for return to service.
BooleanCheck - -
Admin Customizer

Easily configure layout, styles, and preferences for your admin interface.

Select Theme
Default
Material
Modern
SaaS
Flat
Minimal
Color Scheme
Light
Dark
System
Topbar Color
Light
Dark
Gray
Gradient
Sidenav Color
Light
Dark
Gray
Gradient
Image
Sidebar Size
Default
Compact
Condensed
On Hover
On Hover - Show
Offcanvas
Layout Position