| # | Title | Type | Limits | Unit | Report | Required | Actions |
|---|---|---|---|---|---|---|---|
| 1 | Visual Inspection Inspect casing, screen, buttons, speaker, connectors, battery compartment and labels. No damage, corrosion or fluid ingress. |
BooleanCheck | - | - | |||
| 2 | Accessories Present Confirm adult pads present and in date. Confirm paediatric pads/attenuator if supplied. Confirm battery present. |
BooleanCheck | - | - | |||
| 3 | Consumables Status Confirm pads and battery are within service life. Replace via Service Action if required (do not fail unless functional impact). |
BooleanCheck | - | - | |||
| 4 | Battery Serial Number |
TextNote | - | - | |||
| 5 | Battery Expiry Date |
TextNote | - | - | |||
| 6 | Speaker and Prompts Confirm voice prompts and tones are clear and audible. |
BooleanCheck | - | - | |||
| 7 | Service Menu Disconnect pads. Prepare electrical safety analyser and defibrillator analyser. |
Instruction | - | - | |||
| 8 | Confirm correct mahine Date & Time |
BooleanCheck | - | - | |||
| 9 | Machine Pixel Test Run machine pixel test |
BooleanCheck | - | - | |||
| 10 | Power-On Self Test Power on device. Confirm no fault codes and readiness indicator OK. |
BooleanCheck | - | - | |||
| 11 | Configuration: Energy Levels E1: 200, E2: 300, E3: 360 |
BooleanCheck | - | - | |||
| 12 | Configuration: Delete after send Expected: False |
BooleanCheck | - | - | |||
| 13 | Configuration: CPR CPR T1: 120, CPR T2: 120, Initial CPR: Off, Preshock: Off, CPR Prompt: On |
BooleanCheck | - | - | |||
| 14 | Configuration – CPR Time After Shock Verify CPR interval after shock is 120 seconds. |
NumericMeasure | - | - | |||
| 15 | Configuration: Pulse Check Expected: False |
BooleanCheck | - | - | |||
| 16 | Configuration: Manual Mode Expected: Off |
BooleanCheck | - | - | |||
| 17 | Configuration – Stacked Shocks Disabled Verify stacked shocks are OFF (single shock then CPR). |
BooleanCheck | - | - | |||
| 18 | Delivered Energy @ 100J (50Ω) Deliver 100J shock into 50Ω and record delivered energy. |
NumericMeasure | 85 - 115 J | J | |||
| 19 | Delivered Energy @ 360J (50Ω) Deliver 360J shock into 50Ω and record delivered energy. |
NumericMeasure | 306 - 414 J | J | |||
| 20 | Defibrillator Analyser Setup Connect LIFEPAK 1000 to calibrated defibrillator analyser set to 50 ohm load using test cable or pads. |
Instruction | - | - | |||
| 21 | ECG: 60bpm NSR |
NumericMeasure | 56 - 64 bpm | bpm | |||
| 22 | ECG: 120bpm NSR |
NumericMeasure | 116 - 124 bpm | bpm | |||
| 23 | ECG: 180bpm NSR |
NumericMeasure | 176 - 184 bpm | bpm | |||
| 24 | ECG: VF -> Defib Prompt Machine prompts user to connect therapy cable when VF detected in ECG mode |
BooleanCheck | - | - | |||
| 25 | Technician Comments Record advisories, service actions performed, parts replaced, or deviations. |
TextNote | - | - | |||
| 26 | Final Engineering Check Reconnect pads, refit battery, clean device, ensure correct reassembly. |
Instruction | - | - | |||
| 27 | Ready for Clinical Use Confirm device operational, labelled, and safe for return to service. |
BooleanCheck | - | - |