Defibrillator Functional – IEC 60601-2-4
v1.0  |  2025-10-02 10:02Published
Test Steps
# Title Type Limits Unit Report Required Actions
1
Visual Inspection
Inspect casing, leads, electrodes, controls, and labels for integrity.
BooleanCheck - - Locked
2
Protective Earth Continuity
Measure resistance from earth pin to accessible conductive parts.
NumericMeasure - Ω Locked
3
Enclosure Leakage Current (NC)
Measure enclosure leakage in normal condition.
NumericMeasure - µA Locked
4
Enclosure Leakage Current (SFC)
Apply single fault condition per test regime and measure leakage.
NumericMeasure - µA Locked
5
Applied-Part Leakage (NC)
Measure patient/applied-part leakage in normal condition.
NumericMeasure - µA Locked
6
Applied-Part Leakage (SFC)
Simulate single fault condition (e.g., open PE) and measure leakage.
NumericMeasure - µA Locked
7
Insulation Resistance
Measure insulation resistance between mains and applied parts.
NumericMeasure - Locked
8
Delivered Energy @ 200 J
Charge to 200 J and discharge into analyzer; record delivered energy.
NumericMeasure 169 - 231 J J Locked
9
Delivered Energy @ Maximum
Charge to maximum selectable energy (e.g., 360 J) and verify delivered energy.
NumericMeasure 305 - 415 J J Locked
10
Charge Time @ 200 J
Measure time to charge to 200 J.
NumericMeasure - s Locked
11
Charge Time @ Maximum
Measure time to charge to maximum energy (e.g., 360 J).
NumericMeasure - s Locked
12
Charge Cancel / Disarm
Verify charge cancellation and safe disarm into the load.
BooleanCheck - - Locked
13
Sync Function (R-Wave Delay)
Verify synchronisation to R-wave; measure sync delay.
NumericMeasure - ms Locked
14
Energy Accuracy Across Loads
Verify delivered energy remains within tolerance across 25–175 Ω loads.
BooleanCheck - - Locked
15
Maximum Energy Verification
Confirm maximum selectable energy output meets tolerance across specified loads.
BooleanCheck - - Locked
16
Defibrillator Proof Test
Verify device withstands defibrillator proof test pulses as specified.
BooleanCheck - - Locked
17
Pacing Output Current (if equipped)
Set nominal pacing current and measure output.
NumericMeasure - mA Locked
18
Pacing Rate Accuracy (if equipped)
Set pacing rate (e.g., 60 bpm) and verify measured rate.
NumericMeasure - bpm Locked
19
ECG Display Calibration
Verify ECG amplitude calibration using a standard calibration signal.
NumericMeasure - mV Locked
20
Heart Rate Accuracy
Simulate ECG at known rates and verify displayed heart rate.
NumericMeasure - bpm Locked
21
Arrhythmia Detection
Simulate VF/VT/asystole/brady/tachy and verify correct detection/alarms.
BooleanCheck - - Locked
22
Alarm Functionality
Verify audible/visual alarms activate appropriately for critical conditions.
BooleanCheck - - Locked
23
Impedance / Lead-Off Detection
Verify patient impedance checks and lead-off detection operate correctly.
BooleanCheck - - Locked
24
Battery Performance
Verify charge/discharge behaviour and runtime under load per manufacturer spec.
BooleanCheck - - Locked
25
Data Logging / Printer Output
Verify event logs and printed outputs are complete and accurate.
BooleanCheck - - Locked
Admin Customizer

Easily configure layout, styles, and preferences for your admin interface.

Select Theme
Default
Material
Modern
SaaS
Flat
Minimal
Color Scheme
Light
Dark
System
Topbar Color
Light
Dark
Gray
Gradient
Sidenav Color
Light
Dark
Gray
Gradient
Image
Sidebar Size
Default
Compact
Condensed
On Hover
On Hover - Show
Offcanvas
Layout Position