| # | Title | Type | Limits | Unit | Report | Required | Actions |
|---|---|---|---|---|---|---|---|
| 1 | Visual Inspection Inspect casing, leads, electrodes, controls, and labels for integrity. |
BooleanCheck | - | - | Locked | ||
| 2 | Protective Earth Continuity Measure resistance from earth pin to accessible conductive parts. |
NumericMeasure | - | Ω | Locked | ||
| 3 | Enclosure Leakage Current (NC) Measure enclosure leakage in normal condition. |
NumericMeasure | - | µA | Locked | ||
| 4 | Enclosure Leakage Current (SFC) Apply single fault condition per test regime and measure leakage. |
NumericMeasure | - | µA | Locked | ||
| 5 | Applied-Part Leakage (NC) Measure patient/applied-part leakage in normal condition. |
NumericMeasure | - | µA | Locked | ||
| 6 | Applied-Part Leakage (SFC) Simulate single fault condition (e.g., open PE) and measure leakage. |
NumericMeasure | - | µA | Locked | ||
| 7 | Insulation Resistance Measure insulation resistance between mains and applied parts. |
NumericMeasure | - | MΩ | Locked | ||
| 8 | Delivered Energy @ 200 J Charge to 200 J and discharge into analyzer; record delivered energy. |
NumericMeasure | 169 - 231 J | J | Locked | ||
| 9 | Delivered Energy @ Maximum Charge to maximum selectable energy (e.g., 360 J) and verify delivered energy. |
NumericMeasure | 305 - 415 J | J | Locked | ||
| 10 | Charge Time @ 200 J Measure time to charge to 200 J. |
NumericMeasure | - | s | Locked | ||
| 11 | Charge Time @ Maximum Measure time to charge to maximum energy (e.g., 360 J). |
NumericMeasure | - | s | Locked | ||
| 12 | Charge Cancel / Disarm Verify charge cancellation and safe disarm into the load. |
BooleanCheck | - | - | Locked | ||
| 13 | Sync Function (R-Wave Delay) Verify synchronisation to R-wave; measure sync delay. |
NumericMeasure | - | ms | Locked | ||
| 14 | Energy Accuracy Across Loads Verify delivered energy remains within tolerance across 25–175 Ω loads. |
BooleanCheck | - | - | Locked | ||
| 15 | Maximum Energy Verification Confirm maximum selectable energy output meets tolerance across specified loads. |
BooleanCheck | - | - | Locked | ||
| 16 | Defibrillator Proof Test Verify device withstands defibrillator proof test pulses as specified. |
BooleanCheck | - | - | Locked | ||
| 17 | Pacing Output Current (if equipped) Set nominal pacing current and measure output. |
NumericMeasure | - | mA | Locked | ||
| 18 | Pacing Rate Accuracy (if equipped) Set pacing rate (e.g., 60 bpm) and verify measured rate. |
NumericMeasure | - | bpm | Locked | ||
| 19 | ECG Display Calibration Verify ECG amplitude calibration using a standard calibration signal. |
NumericMeasure | - | mV | Locked | ||
| 20 | Heart Rate Accuracy Simulate ECG at known rates and verify displayed heart rate. |
NumericMeasure | - | bpm | Locked | ||
| 21 | Arrhythmia Detection Simulate VF/VT/asystole/brady/tachy and verify correct detection/alarms. |
BooleanCheck | - | - | Locked | ||
| 22 | Alarm Functionality Verify audible/visual alarms activate appropriately for critical conditions. |
BooleanCheck | - | - | Locked | ||
| 23 | Impedance / Lead-Off Detection Verify patient impedance checks and lead-off detection operate correctly. |
BooleanCheck | - | - | Locked | ||
| 24 | Battery Performance Verify charge/discharge behaviour and runtime under load per manufacturer spec. |
BooleanCheck | - | - | Locked | ||
| 25 | Data Logging / Printer Output Verify event logs and printed outputs are complete and accurate. |
BooleanCheck | - | - | Locked |